Products
Plants
Ecosystem
Soil
Light
Water
Gas Monitoring
Contact
Contact LI-COR
Press & Media
Corporate
International Distributors
LICORbio
Solutions
Greenhouse Gas Monitoring
Agricultural Best Practices
Environmental Protection
Soil, Atmosphere, & Ocean Studies
Energy & Indoor Environmental Monitoring
Water Monitoring
Resources
Support Site
Publications
Webinars
Conferences
News
Press & Media
Blog
Products
All Products
Plants
Ecosystem
Soil
Light
Water
Gas Monitoring
Solutions
Greenhouse Gas Monitoring
Agricultural Best Practices
Environmental Protection
Energy & Indoor Environmental Monitoring
Soil, Atmosphere, & Ocean Studies
Water Monitoring
Resources
Support
Webinars
Publications
Conferences
Professional Services
Contact
LI-COR
International Distributors
Press & Media
Corporate
LICORbio
News
Blog
Press & Media
cart script
LI-830 and LI-850 Gas Analyzers
Overview
Integrations
Specifications
Resources
Get a Quote
Specifications
CO
2
measurements
Measurement range:
0 to 20,000 ppm
Accuracy
LI-850
Within 1.5% of reading
LI-830
Within 3% of reading
Calibration drift
Zero Drift
1
:
<0.15 ppm / °C
Span Drift
2
:
<0.03% / °C
Total drift at 370 ppm:
<0.4 ppm / °C
RMS noise at 370 ppm with 1 sec signal filtering:
<1 ppm
Sensitivity to water vapor (
LI-850
only):
<0.1 ppm CO
2
/ mmol mol
-1
H
2
O
Lower limit of detection
1.5 ppm
H
2
O
measurements (
LI-850
only)
Measurement range:
0 to 60 mmol mol
-1
Accuracy:
Better than 1.5% of reading
Calibration drift
Drift at 0 mmol mol
-1
:
<0.005 mmol mol
-1
/ °C
Span Drift
1
at 10 mmol mol
-1
:
<0.006 mmol mol
-1
/ °C
Total Drift at 10 mmol mol
-1
<0.016 mmol mol
-1
/ °C
RMS noise at 10 mmol mol with 1 sec signal filtering:
<0.01 mmol mol
-1
Sensitivity to CO
2
:
<0.0001 mmol mol
-1
H
2
O / ppm CO
2
Lower limit of detection:
<0.015 mmol mol
-1
General
Output rate:
Up to 2 measurements per second
Response time (T90)
CO
2
:
<3.5 seconds from 0 to 375 ppm
H
2
O:
<3.5 seconds from 0 to 21 mmol mol
-1
Internal optical cell volume:
14.5 mL
Measurement principle:
Non-Dispersive Infrared
Traceability:
CO
2
:
Traceable gases to WMO standards from 0 to 3,000 ppm; traceable gases to EPA protocol gases from 3,000 to 20,000 ppm
H
2
O (
LI-850
only):
NIST traceable
LI-610
Portable Dew Point Generator
Pressure compensation range:
50 to 110 kPa
Maximum gas flow rate:
1 liter min
-1
Output signals:
Two analog voltage (0 to 2.5 V or 0 to 5 V) and two current (4 to 20 mA)
Digital outputs:
TTL (0 to 5 V) or Open Collector
DAC resolution:
16 bits across user-specified range
Power requirements
Input voltage:
12-30 VDC
After warmup (without pump):
0.33A @ 12 VDC (4.0 W) average
After warmup (with pump):
0.42A @ 12 VDC (5.0 W) average
During warmup:
1.2 A @ 12 VDC (14 W) maximum
Operating temperature range:
-20°C to 45 °C
Relative humidity range:
0 to 95% RH, Non-condensing
Dimensions:
22.23 cm W x 15.25 cm D x 7.62 cm H
Weight
No pump, no display:
1.0 kg
No pump, with display:
1.02 kg
With pump, no display:
1.3 kg
With pump, with display:
1.32 kg
Pump specifications (optional)
Operating temperature range:
5 to 45 °C
Storage temperature range:
-20 to 60 °C
Operating humidity range:
0 to 80% RH
Nominal flow rate:
0.75 liters minute
-1
Power consumption:
1 W (nominally)
Expected life span:
8,000 hrs in standard conditions with a normal load
Display specifications (optional)
Dimensions:
6.7 cm corner-to-corner
Resolution:
400 x 200 px; monochrome
Power consumption:
<200 µW
Displayed variables:
CO
2
reading,
H
2
O
reading (
LI-850
only), optical bench temperature, and pressure.
Specifications subject to change without notice.